DVD1260 Viscosimeter

The SECcurity DVD1260 four capillary viscometer lowers the detection limit with an asymmetric bridge design. While other viscometers dilute 50% of the sample in the solvent path, the DVD1260 dilutes only 20% of the sample in the solvent path. 80% of the samples run through the sample path leading to an unparalleled signal/noise ratio without any digital signal processing.

Low-noise electronics combined with high-sensitivity pressure transducers results in excellent peak shapes and high signal linearity. The use of reservoirs (instead of hold-up columns) eliminates viscometer modifications when the column setup changes and reduces the analysis time to 50% compared with other viscometers. 

El viscosímetro DVD1260 puede ser integrado en cualquier GPC/SEC o sistema FFF.

 

  Especificaciones

 Tipo de detección

 viscosímetro de 4 capilares en línea

 Señales

 Simultánea medición de presión delta y presión de entrada

 Nivel de ruido

 0.2 Pa, diferente Canal de presión, 25°C

 Celda

 Puente asimétrico con división 80/20

 Transductores de presión

 Rápida respuesta Hastelloy transductores de presión sin histéresis.

 Rango de Temperatura

 Ambiente hasta 80°C

 Estabilidad Térmica

 > 0.01°C

 Precisión de Temperatura

 ± 0.5°C

 Caudal máximo en puente

 1.5mL/min analítico, bioinerte
 0.5mL/min micro

 Entrada analógica

 2 ins analógicos para cualquier detector de concentración (Inicio, Error, Flush, Zero)

 Salida analógica

 1.0 Volt/10 kPa FSD Presión Diferencial
 1.0 Volt/100 kPa FSD Presión de entrada

 GLP/Cumplimiento

 Libro de registro de errores internos, protección transductor de presión electrónico con código de error.