Electrolytic Conductivity Detector (ELCD)
The Model 5320 Electrolytic Conductivity Detector (ELCD) is a third-generation ELCD designed for selective detection of halogen-containing compounds. The 5320 ELCD consists of three principal components: the reactor assembly, the cell-solvent assembly, and the detector controller.
The ELCD’s principal mode of operation is the halogen mode (X); sulfur (S) and nitrogen (N) modes are also available. Each detection mode kit contains all of the required materials except the solvent) to operate the ELCD in the specified mode.
- Diseño de cambio rápido del reactor, cartucho de resina desechable y sistema de solvente fiable.
- Controlador analógico de la temperatura y del flujo de solvente del reactor.
- Base del detector optimizado para columnas capilares.
- Ventilación mediante GC, programación del tiempo en relé.
- Interfase directa con la mayoría de marcas y modelos de GC.
- Interfase directa con el Modelo 4430 PID sin línea de transferencia; el diseño tándem ocupa sólo un puerto de detector.
Detección de masas
|
|
Máximo
|
1 pg. Lindano
|
Máximo
|
5 µg. Lindano
|
Rango dinámico
|
5 x 106
|
Selectividad
|
CI/HC > 106
CI/N > 105
CI/S > 105
|
Temperatura del reactor
|
|
Rango
|
800ºC – 1.100ºC en incrementos de 100ºC
|
Estabilidad
|
+/- 1ºC
|
Flujo solvente
|
Ajustable a la ampliación de la celda
|
Rango de flujo de solvente
|
0-200µL/min
|
Válvula de venteo
|
Controlada por un GC,
|
Detector de salida
|
0-1V o 0-10V
|
Modos operativos
|
Halógeno, Azufre, Nitrógeno
|
Requisitos Gas
|
Hidrógeno, Pureza Ultra Alta (99.999% o mayor)
|
Requisitos de potencia
|
90-260 VAC (+/- 10%)
47-63Hz, 200W
|
Controlador del detector
|
|
Peso
|
3.8Kg (8.4 lb)
|
Dimensiones
|
21.0cm H x 12.7cm W x 30.5cm D (8.25” H x 5.0” W x 12” D)
|