NG POLLUCE
POLLUCE is designed to satisfy the needs of all LCMS brand on the market. It makes use of Pressure Swing Adsorption technology (PSA), through the use of specific selected molecular sieves (CMS), enable to obtain extremely pure nitrogen, this technologies reduced risk of gas contamination and phthalates-free.
The generator could be equipped with a built-in oil less compressor with inverter technology. Inverter electronically control the electrical voltage, current and frequency of the compressor motor. It receives information from sensors monitoring operating conditions, and adjust the revolution speed of the compressor, which directly regulates air pressure output. Optimum control of operation frequency results in eliminating the consumption of excessive electricity and providing the most comfortable conditions to increase the compressor life-time. Thanks to the CPU with touch screen display, the continuous monitoring of operational status is guaranteed.
Main application
- LCMS
- Solvent evaporation
Ventajas
-
Tecnología PSA
-
Fácil uso e instalación (plug and play)
-
Bajo nivel de ruido
-
Genera nitrógeno bajo demanda con control de flujo electrónico exclusivo
-
Compresor de baja energía exclusivamente controlado por la tecnología inverter
-
Diseñado para funcionar las 24 horas del día
-
Presión hasta 8 bares
-
Opción de depurador de horno CH4
Modelos: NG POLLUCE |
40 |
40C |
100 |
Salida N2 |
|||
Flujos (máx.) |
40 L/min |
40 l/min |
100 L/min |
Presión de salida (Máx.) |
Presión de entrada –2 bares |
Hasta 8 bares |
Presión de entrada –2 bares |
Pureza de Nitrógeno |
Hasta 99% |
||
Salida de Punto de rocío |
<-50°C |
||
Salida de partículas |
0.01 micrones |
||
Entrada de aire |
|||
Requisito de entrada de aire |
Partícula: <1 micron |
||
Punto de rocío: < -20°C (- 4°F) |
|||
Aceite: <0.01 mg/m³ |
|||
Flujo (Máx.) |
100 L/min |
250 L/min |
|
Presión de suministro (min–máx.) |
7 bares (95 psi) / 10 bares (145 psi) |
||
Temperatura Recomendada |
< 30°C (86°F) |
||
Comunicación |
|||
LCD pantalla táctil |
Estándar |
||
RS485 |
Estándar |
||
RS232 |
Estándar |
||
USB |
Estándar |
||
Digital I/O |
Estándar |
||
Información General |
|||
Voltaje de suministro |
100-240Vac 50/60Hz |
Monofásico 220-40Vac 50/60Hz |
100-240Vac 50/60Hz |
Tipo de conexión |
IEC320-C13 |
Cable, 3x2.5mm² |
IEC320-C13 |
Potencia de instalación (máx.) |
280W |
2.2KW |
280W |
Capacidad del fusible (5x20mm) |
4A - 250VAC |
- |
4A - 250VAC |
Peso neto |
<160Kg |
<300Kg |
<220Kg |
Dimensiones (W x D x H) |
42 x 40 x 160 cm |
42 x 90 x 160 cm |
42 x 80 x 140 cm |
Conexiones |
|||
Salida |
¼" BSPP hembra |
||
Entrada |
¼" BSPP hembra |
- |
¼" BSPP hembra |
Purga de agua |
Tubo de 8mm fácil de ajustar |
¼" BSPP hembra |
Tubo de 8mm fácil de ajustar |
Condiciones de funcionamiento/ almacenamiento |
|||
Temperatura de funcionamiento |
5-40°C (41-104°F) |
||
Temperatura de almacenamiento |
1-50°C (34-122°F) |
||
Humedad (máximo, sin condensación) |
70% |
||
Ruido |
<25dB(A) |
<65dB(A) |
<25dB(A) |
IP |
IP20 |
||
Grado de contaminación |
2 (sin compuestos aromáticos) |
||
Altitud |
< 2000m |
Códigos
Gas Generadores |
6920.60.040 NG POLLUCE 40 |
6920.60.040C NG POLLUCE 40C |
|
6920.60.100 NG POLLUCE 100 |
|
Opciones |
6920.61.001 CH4 depurador para NG POLLUCE (solo para 40/ 40C) |
Consumibles |
6920.61.002 Kit de servicio de 8000 horas para NG POLLUCE |