UNITY-xr Markes International

The most versatile and powerful thermal desorption unit.

UNITY-xr™ is an industry-leading analytical​ TD instrument for the simultaneous analysis of VOCs, SVOCs and thermally labile compounds collected onto sorbent tubes. 

In contrast to other thermal desorption units, UNITY-xr quantitatively re-collects samples for re-analysis or storage, enhancing sample security. UNITY-xr also features an electrically cooled cold trap, dispensing with the expense and inconvenience of liquid cryogen.

UNITY-xr is perfect for any laboratory needing a versatile thermal desorber for all their TD applications.

Product benefits:

  • With quantitative sample re-collection for repeat analysis, UNITY-xr offers simple method/data validation and overcomes the ‘one-shot’ limitation of other systems.
  • Analyse the widest possible range of VOCs and SVOCs in a single run, including reactive and thermally labile species.
  • Reduce running costs by avoiding the need for cryogen.
  • Enhance productivity with the world’s most versatile thermal desorber – UNITY-xr offers unmatched analytical performance for the widest range of TD applications.
  • Additional modules allow automation of multiple tubes, canisters and on-line samples – making UNITY-xr the perfect platform for expansion as the needs of your lab grow.
  • You’re using a desorber that complies with key standards such as US EPA Method TO-17.

Features:

  • Quantitative sample re-collection of all the split flows enables repeat analysis of critical samples and easy method validation.
  • An inert, optimised flow path allows quantitative recovery of C2 to C44, including reactive and thermally labile species... from percent to sub-ppt concentrations.
  • As well analysing 3½″ standard TD tubes, the UNITY-xr facilitates analysis of on-line, canister and bag samples through the modular addition of versatile accessories.
  • Rapid operation of the electrically-cooled focusing trap reduces costs and ensures fast sample throughput.
  • Method-compliance is aided by tube- and leak-testing, water management, and addition of internal standards.
  • Samples are easily tracked with barcoded tubes and RFID TubeTAGs.
  • Compatible with all major makes of GC and GC–MS.
Desorption tubes
 The automated desorber must be compatible with industry standard 3.5-inch (89 mm) x ¼-inch (6.4 mm)  O.D. tubes for compliance with standard methods e.g. ASTM D 6196, EN ISO 16017, ISO 16000-6, EN  14662-4, US EPA Method TO-17, NIOSH 2549 and the UK MDHS series.
 The automated desorber must be compatible with RFID tagged tubes
 The tubes must be compatible with standard Swagelok-type ¼-inch long-term brass storage caps fitted with  combined PTFE ferrules as specified in standard methods.
 Sample tubes must be compatible with axial diffusive sampling as per standard methods (EN ISO 16017-2,  EN 14662-4 and ASTM D 6196).
 Tubes should be orientated horizontally throughout TD operation to prevent particles dropping into tube  seals and to prevent samples shifting within the tube during direct desorption of materials.
 Carrier gas must only be allowed to pass through the middle of the tube and not around the outside of the  tube during the analysis process – to minimise contamination. 

 

Focusing trap
 The system must feature 2-stage desorption using an electrically-cooled packed cold trap capable of cooling  to sub-zero temperatures as specified in US EPA Method TO-17 and other standards.
 It should be possible to pack the electrically-cooled cold trap with up to four sorbent beds to a total length of  60 mm to match the sorbent bed length found in industry standard sample tubes. 
 Desorption of the focusing trap must take place in a backflush mode i.e. analytes must enter the trap from  one end and be desorbed from the same end to facilitate simultaneous analysis of volatiles and semi-  volatiles. 
 The system should offer rapid heating (>60°C/sec) of the focusing trap to optimise chromatographic  resolution and analytical sensitivity. 
 Optional, slower, programmed heating options should also be available for the focusing trap to ensure  compatibility with very reactive components
 The focusing trap must feature a narrow / restricted I.D. (<1 mm I.D.) at the inlet / outlet end to the cold  trap to optimise linear gas velocity and minimise band width after trap desorption.
 The dimensions, sorbent mass and heating rate of the focusing trap plus the design of the sample flow path  downstream of the focusing trap should facilitate splitless operation even with high resolution  capillary GC / GC-MS operating at flow rates of 2 mL/min or below.
 The flow of dry gas required for purging the cold trap area to prevent build-up of ice during system  operation, should be less than 100 ml/min to minimise gas consumption and operator intervention.

 

Sample Flow Path
 The desorber must feature a totally inert flow path (constructed, for example, using SilcosteelÒ, quartz or  fused silica components) the flow path must contain no uncoated metal fittings.
 The sample flow path inside the desorber must feature uniform heating and minimal volume / length / I.D.,  particularly post focusing trap, to ensure high linear gas velocity and no band dispersion as components are  transferred to the capillary column or fused silica retention gap.
 It must be possible to set the flow path at temperatures below 100°C to prevent degradation of heat  sensitive compounds.  It must also be possible to set the flow path at temperatures which allow quantitative  recovery of low volatility compounds, including n-C40
 After primary (tube) desorption, the desorber must isolate the cooling sample tube from the flow path to  prevent contaminants desorbing late from the primary tube from passing into the focusing trap after  secondary (trap) desorption has begun.  Isolation of the primary tube should involve some sort of  appropriate heated, inert valve
 At no stage during sample analysis must the sample flow path be opened to atmosphere or air (and  associated contaminants) will get into the system
 Once the desorber has initiated a run, all parts of the desorption sequence – leak testing, purging, tube  desorption, trap desorption and starting the GC run – should be automatic, to allow unattended operation.
 The desorber must connect directly to the analytical column or to a heated uncoated, deactivated fused-silica  capillary retention gap, without passing through a GC injector

 

Sample split capabilities
 It should be possible to split both during primary (tube) desorption (inlet splitting) and secondary (trap)  desorption (outlet splitting) – in order to obtain overall split ratios in excess of 1000:1 for analysis of high  concentration samples.
 The flow path to the sample split point(s) must be inert and uniformly heated
 The system must allow the user to select for the sample split flow to be on during instrument standby to  minimise air ingress for MS and ECD installations.
 The system must allow manual, quantitative re-collection of both the inlet and outlet split effluent on a  single re-collection tube such that the entire unanalysed portion of the sample, from any single or double  split method, is re-collected. The re-collected sample is needed for repeat analysis, sample archiving and  method / data validation in compliance with ASTM 6196.
 There must be easy access to the split re-collection tube without need to tighten and undo screw fittings.  Software must also allow carrier gas to be isolated from the split line such that access to the re-collection  tube does not cause a temporary leak.
 It must be possible to automate split re-collection – this upgrade should be possible at any point during the  lifetime of the system

 

Analytical sequence
 As required by relevant standard methods such as US EPA Method TO-17, the analytical sequence must  include an ambient temperature, no-flow leak test of each tube prior to analysis, to ensure data integrity.
 Any tube which fails the leak test must remain intact and unanalysed, maintaining sample integrity until the  fault is diagnosed and the sample can be successfully analysed.
 The analytical sequence must also include an ambient temperature carrier gas purge of each tube to  eliminate O2 prior to analysis.  Any air purged from the system must be directed away from the analytical  GC(MS) system. This is in order to minimise interference – e.g. column oxidation or high air / water  background on MS.
 The system must offer the option of an elevated temperature tube purge during standard 2-stage tube  desorption for selective elimination of water and unwanted volatiles prior to analysis.
 The desorption oven should heat from near ambient during primary (tube) desorption to minimise risk of  split discrimination from samples with high solvent or water content
 The system should facilitate the sequencing of multiple desorptions of a single tube at different temperatures  for automating the TD method development process

TD40.Direct desorption of VOCs and SVOCs from leather furnishings.pdf TD33.Analysis of the interior atmosphere of a passenger car by TD-GC-MS.pdf TD14.Monitoring labile, high-boiling organic vapours such as those found in -clean room air- in semi-conductor fabrication plants.pdf TD28.Optimising Analytical Performance and Extending the application range of thermal desorption for monitoring air indoors and inside vehicle cabins.pdf TD56.TD-GC-MS analysis of VOCs for material emissions testing.pdf TD55.Using the FLEC to determine emissions of VOCs from materials and products.pdf TD57.Characterisation of paint samples by direct desorption TD-GC-MS.pdf TD59.Direct desorption of car trim materials for VOC and SVOC analysis in avvordance whit VDA method 278.pdf TD62.Materials emissions testing in the semiconductor and electronics industries.pdf TD65.Automating the measurement of VOCs ans SVOCs in materials using direct thermal desorption.pdf TD66.Improving the identification and measurement of trace odorous and toxic compnents during material emissions testing.pdf TD67.Introducing the Micro-Chamber-Thermal Extractor for rapuid emissions screening of products and materials.pdf TD70.Materials emissions testing using the FLEC® Cell, testing recovery and sink effects.pdf TD69.Correlation of emission rates determined using the Micro.Chamber-Thermal Extractor whit those for conventional emission-chamber tests.pdf TD68.Using Markes emission screening technology to simplicity compliance whit construction product regulations.pdf TD71.Summary of FLEC Applications.pdf TD72.An Introduction to Emission Cells and comparison to small chambers for material emissions testing.pdf TD73.The performance of Markes Micro-chamber- Thermal Extractor for low-cost, fast and meaningful material emissions screening.pdf TD89.Thermal desorption technology for testing chemical emissions from construction products and consumer goods.pdf TD90.Automated detection of trace target compounds in complex chemical emission profiles from products and materials.pdf TD93.Rapid microchamber tests for screening chemical emissiond from car trim in accordance with ISO 12219-3.pdf TD74.The suitability of emission cells for emissions testing and a comparison with samll chambers.pdf TD103.Enhancing the development of low-emitting products and materials.pdf

TD3.National and international standard methods relating to speciated monitoring of vapour-phase organic compounds in air.pdf TD5.Advice on sorbent selection, tube conditioning, tube storage and air sampling.pdf TD15.On-line process control of speciated organic chemicals in industrial gases at sub-ppb levels.pdf TD18.Developments in the determination of nitrous oxide using TD-GC.pdf TD16.Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air.pdf TD32.The analysis of sulfur compounds using on-line and off-line TD-GC.pdf TD29.VOC air monitoring technology and its application to contamined land.pdf TD35.Thermal desorption of dioctyl phthalate and other plasticisers.pdf TD27.A review of sorbent-based sampling methods for VOCs and SVOCs in air.pdf TD38.Occupational Exposure Limit Levels for VOCs compatible whit TD-GC.pdf TD37.Workplace Air Monitoring using Pumped sampling onto sorbent tubes and analysis by TD-GC-MS.pdf TD46B.Monitoring organic vapours in air – A comparison of thermal desorption and carbon disulfide(CS2) extraction.pdf TD46.Monitoring organic vapours in air – A comparison of.pdf TD53.Quantitative recovery and method validation of high-boiling SVOCs using TD-GC-MS.pdf TD47.The Analysis of Landfill Gas Compounds using TD-GC-MS.pdf TD49Fenceline and ambient monitoring of benzene and other hydrocarbons using diffusive samping and TD analysis.pdf TD77.Using Thermal Desorption for Industrial (Stack) emissions testing.pdf TD81.Analysis of canister air samples using cryogenfree thermal desorption in compliance with US EPA method TO-15.pdf TD80.Evaluation of a ‘Soil Gas’ sorbent tube for improving the measurement of volatile and semivolatile fuel vapors in contamined lan.pdf TD86.Monitoring 'air toxics' in ambient air using sorbent tubes by automated, cryogen-free thermal desorption on accordance with US EPA method TO-17.pdf TD87.A cryogen-free method for monitoring trace greenhouse gases in air.pdf TD97.Analysis of polycyclic aromatic hydrocarbons from vehicle exhaust using TD-GC-MS.pdf TD99.Using the CIA Advantage for automated cryogenfree analysis of canister air and gas.pdf TD102.Investigation of the performance of sorptive extraction cartbrige for air monitoring.pdf

TD2.Prediction of uptake rates for diffusive tubes.pdf TD1.Uptake rates for tube-type axial diffusive samplers.pdf TD6.Common problems with thermal desorption and how to solve them.pdf TD4.Publications and presentations citing Markes’ products.pdf TD7.Calibration, Preparing and introducing thermal desorption standards using sorbent tubes.pdf TD8.Principles of diffusive monitoring.pdf TD10.Use of diffusive samplers with TD–GC analysis for monitoring VOCs in ambient air.pdf TD9.Monitoring materials and processes for VOCs at high and trace levels.pdf TD17.The GC-FID sensitivity of the UNITY thermal desorber.pdf TD13.Evaluation of the Bio-VOC- a low-cost, simple device for biological monitoring of VOCs in breath.pdf TD19.Minimising artefacts – Considerations for storage and transport of sorbent tubes.pdf TD20.Confirming sorbent tube retention volumes and checking for analyte breakthrough.pdf TD21.Developing and optimising TD methods.pdf TD22.Selection of gas flows and split ratios during thermal desorption.pdf TD24.Overcoming the ‘one-shot’ limitation of thermal desorption by the re-collection of desobed samples.pdf TD25.Calculating atmospheric concentrations from analyte masses retained on sorbent tubes.pdf TD26.Minimising analytical interference from water during the analysis of sorbent tubes.pdf TD30.Certified Reference Materials for Analysis of VOCs in air by TD-GC.pdf TD35.Thermal desorption of dioctyl phthalate and other plasticisers.pdf TD12.Analytical thermal desorption, History, technical aspects and application range.pdf TD60.The complementary role of thermal desorption and other sample introduction technologies fo GC.pdf TD42.Uptake rates for radial diffusive samplers.pdf TD61.Diffusion-locking technology.pdf TD75.Liquid standard injection, tube impedance and other factors that may cause discrimination during the calibration of TD methods.pdf TD64.Simultaneous TD-GC(-MS) analysis of VOCs and SVOCs.pdf TD79.Air monitoring - the respective advantages and applications of canisters and tubes.pdf TD85.GC-MS data analysis using ClearView™ background compensation software,dealing with common ions.pdf TD82.TubeTAG™ - Enhanced tracking of sampleand and tube-related infromation for thermal desorption.pdf TD83.Using ClearView background compensation software to enhance GC-MS methods for trace compound analysis.pdf